5
Specifications
166
E4418B Power Meter User’s Guide
For Agilent E-Series power sensors
In FAST mode, within the range –50 dBm to +17 dBm, for a 10 dB decreasing
power step, the settling time is 10 ms
1
.
1
When a decreasing power step crosses the power sensor’s auto-range switch point, add 25 ms. Refer to the
relevant power sensor manual for further information.
For Agilent E- Series power sensors in normal and x2 speed modes, manual
filter, 10 dB decreasing power step:
Table 5-5
Settling Time
Number of Averages
1
2
4
8
16
32
64
128
256
512
1024
Settling Time (s)
(Normal Mode)
0.07
0.12
0.21
0.4
1
1.8
3.3
6.5
13
27
57
Response Time (s)
(x2 mode)
0.04
0.07
0.12
0.21
0.4
1
1.8
3.4
6.8
14.2
32
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...