4
Error Messages
150
E4418B Power Meter User’s Guide
–178
Expression data not allowed
A legal expression data was encountered but not allowed by the power meter at this point.
For example
SYST:LANG (5+2)
.
–211
Trigger ignored
Indicates that
<GET>
or
*TRG
or
TRIG:IMM
was received and recognized by the device
but was ignored because the power meter was not in the wait for trigger state.
–213
Init ignored
Indicates that a request for a measurement initiation was ignored as the power meter was
already initiated.
For example,
INIT:CONT ON
INIT
.
–214
Trigger deadlock
TRIG:SOUR
was set to
HOLD
or
BUS
and a
READ?
or
MEASure?
was attempted,
expecting
TRIG:SOUR
to be set to
IMMediate
.
–220
Parameter error;Frequency list must be in ascending order.
Indicates that the frequencies entered using the
MEMory:TABLe:FREQuency
command are not in ascending order.
–221
Settings conflict
This command occurs under a variety of conflicting conditions. The following list gives a few
examples of where this error may occur:
•If the
READ?
parameters do not match the current settings.
•If you are in fast mode and attempting to switch on for example, averaging, duty cycle or
limits.
•Trying to clear a sensor calibration table when none is selected.
–221
Settings conflict;DTR/DSR not available on RS422
DTR/DSR is only available on the RS232 interface.
–222
Data out of range
A numeric parameter value is outside the valid range for the command. For example,
SENS:FREQ 2KHZ
.
–224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the command. You may
have used an invalid parameter choice.
For example,
TRIG:SOUR EXT
.
–226
Lists not same length
This occurs when
SENSe:CORRection:CSET[1]|CSET2:STATe
is set to
ON
and the frequency and calibration/offset lists do not correspond in length.
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...