5
Specifications
162
E4418B Power Meter User’s Guide
Measurement Noise
Power sensor dependent (refer to
and
).
Averaging effects on measurement noise. Averaging over 1 to 1024
readings is available for reducing noise.
provides the
measurement noise for a particular power sensor with the number of
averages set to 16 for normal mode and 32 for x2 mode. Use the “Noise
Multiplier” for the appropriate mode (normal or x2) and number of
averages to determine the total measurement noise value.
For example, for an Agilent 8481D power sensor in normal mode with the
number of averages set to 4, the measurement noise is equal to:
(<45 pW x 2.75) = <124 pW
Table 5-2
Noise Multiplier
Table 5-3
Power Sensor Specifications±
Number of Averages
1
2
4
8
16
32
64
128
256
512
1024
Noise Multiplier
(Normal Mode)
5.5
3.89
2.75
1.94
1.0
0.85
0.61
0.49
0.34
0.24
0.17
Noise Multiplier
(x2 mode)
6.5
4.6
3.25
2.3
1.63
1.0
0.72
0.57
0.41
0.29
0.2
Power Sensor
Zero Drift
1
Measurement Noise
2
Agilent 8481A
3
<±10 nW
<110 nW
Agilent 8481B
<±10 mW
<110 mW
Agilent 8481D
<±4 pW
<45 pW
Agilent 8481H
<±1 mW
<10 mW
Agilent 8482A
<±10 nW
<110 nW
Agilent 8482B
<±10 mW
<110 mW
Agilent 8482H
<±1 mW
<10 mW
Agilent 8483A
<±10 nW
<110 nW
Agilent 8485A
<±10 nW
<110 nW
Agilent 8485D
<±4 pW
<45 pW
Agilent R8486A
<±10 nW
<110 nW
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...