![Agilent Technologies E4418B Скачать руководство пользователя страница 192](http://html.mh-extra.com/html/agilent-technologies/e4418b/e4418b_user-manual_2868035192.webp)
5
Specifications
168
E4418B Power Meter User’s Guide
For Agilent N8480 Series power sensors
Manual filter, 10 dB decreasing power step:
Table 5-6
Settling Time
Auto filter, default resolution, 10 dB decreasing power step,
normal and x2 speed modes:
Number of Averages
1
2
4
8
16
32
64
128
256
512
1024
Settling Time (s)
(Normal Mode)
0.15
0.2
0.3
0.5
1.1
1.9
3.4
6.6
13
27
57
Response Time (s)
(x2 mode)
0.15
0.18
0.22
0.35
0.55
1.1
1.9
3.5
6.9
14.5
33
10 dB
10 dB
10 dB
20 dB
Minimum dBm
Power
150 ms
6.6 s
1 s
200 ms
Settling
Times
Sensor
Dynamic
Range
Maximum dBm
Normal Mode
Typical
13.5 s
5 dB
10 dB
10 dB
10 dB
20 dB
Power
150 ms
3.6 s
400 ms
180 ms
Settling
Times
Sensor
Dynamic
Range
Maximum dBm
Minimum dBm
x2 Mode
Typical
6.6 s
5 dB
Содержание E4418B
Страница 1: ...Agilent Technologies Agilent E4418B Power Meter User s Guide ...
Страница 14: ...xiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...xx E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...xxii E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 24: ...xxiv E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 47: ...Power Meter Operation 2 E4418B Power Meter User s Guide 23 Contacting Agilent Technologies on page 103 ...
Страница 113: ...Power Meter Operation 2 E4418B Power Meter User s Guide 89 Figure 2 21 Save Recall Screen ...
Страница 126: ...2 Power Meter Operation 102 E4418B Power Meter User s Guide Figure 2 23 Replacing the fuse In line fuse Spare fuse ...
Страница 168: ...3 Menu Reference 144 E4418B Power Meter User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 169: ...145 E4418B Power Meter User s Guide Agilent Technologies 4 Error Messages Introduction 146 Error Messages 148 ...