Chapter 5 Error Messages
Execution Errors
119
5
-430
Query DEADLOCKED
A command was received which generates too much data to fit in the output
buffer and the input buffer is also full. Command execution continues but
all data is lost.
-440
Query UNTERMINATED after indefinite response
The
*IDN?
command must be the last query command within a command
string.
Example:
*IDN?;:SYST:VERS?
501
Isolator UART framing error
502
Isolator UART overrun error
511
RS-232 framing error
512
RS-232 overrun error
513
RS-232 parity error
514
Command allowed only with RS-232
There are three commands which are only allowed with the RS-232
interface:
SYSTem:LOCal
,
SYSTem:REMote
, and
SYSTem:RWLock
.
521
Input buffer overflow
522
Output buffer overflow
550
Command not allowed in local
You should always execute the
SYSTem:REMote
command before sending
other commands over the RS-232 interface.
800
P25V and N25V coupled by track system
The
OUTP:TRAC
should be off when coupling between the +25V output and
the -25V output.
801
P25V and N25V coupled by trigger subsystem
The +25V output and the -25V output should be uncoupled to enable the
tracking operation for those outputs.
Содержание E3631-90002
Страница 9: ...8...
Страница 14: ...1 1 General Information...
Страница 25: ...24...
Страница 26: ...2 2 Initial Operation...
Страница 34: ...3 3 Front Panel Operation...
Страница 64: ...4 4 Remote Interface Reference...
Страница 113: ...112...
Страница 114: ...5 5 Error Messages...
Страница 124: ...6 6 Application Programs...
Страница 137: ...136...
Страница 138: ...7 7 Tutorial...
Страница 153: ...152...
Страница 154: ...8 8 Specifications...
Страница 160: ...Chapter 8 Specifications Supplemental Characteristics 159 8 Figure 8 1 Dimensions of Agilent E3631A Power Supply...
Страница 161: ...160...