Chapter 4 Remote Interface Reference
The SCPI Status Registers
97
4
To Determine When a Command Sequence is Completed
1
Send a device clear message to clear the power supply's output buffer (e.g.,
CLEAR 705
).
2
Clear the event registers with the
*CLS
(clear status) command.
3
Enable the “operation complete” bit (bit 0) in the Standard Event register by
executing the
*ESE 1
command.
4
Send the
*OPC?
(operation complete query) command and enter the result
to ensure synchronization.
5
Execute your command string to program the desired configuration,
and then execute the
*OPC
(operation complete) command as the last
command. When the command sequence is completed, the “operation
complete” bit (bit 0) is set in the Standard Event register.
6
Use a serial poll to check to see when bit 5 (standard event) is set in the
Status Byte summary register. You could also configure the power supply
for an SRQ interrupt by sending
*SRE 32
(Status Byte enable register, bit 5).
Using *OPC to Signal When Data is in the Output Buffer
Generally, it is best to use the “operation complete” bit (bit 0) in the
Standard Event register to signal when a command sequence is completed.
This bit is set in the register after an
*OPC
command has been executed. If
you send
*OPC
after a command which loads a message in the power
supply's output buffer (query data), you can use the “operation complete” bit
to determine when the message is available. However, if too many messages
are generated before the
*OPC
command executes (sequentially), the output
buffer will fill and the power supply will stop processing commands.
Содержание E3631-90002
Страница 9: ...8...
Страница 14: ...1 1 General Information...
Страница 25: ...24...
Страница 26: ...2 2 Initial Operation...
Страница 34: ...3 3 Front Panel Operation...
Страница 64: ...4 4 Remote Interface Reference...
Страница 113: ...112...
Страница 114: ...5 5 Error Messages...
Страница 124: ...6 6 Application Programs...
Страница 137: ...136...
Страница 138: ...7 7 Tutorial...
Страница 153: ...152...
Страница 154: ...8 8 Specifications...
Страница 160: ...Chapter 8 Specifications Supplemental Characteristics 159 8 Figure 8 1 Dimensions of Agilent E3631A Power Supply...
Страница 161: ...160...