Chapter 3 Front-Panel Operation
GPIB Interface Configuration
53
3
GPIB Interface Configuration
The GPIB connector on the rear panel connects your power supply to the
computer and other GPIB devices. Chapter 1 lists the cables that are
available from Agilent Technologies. An GPIB system can be connected
together in any configuration (star, linear, or both) as long as the following
rules are observed:
• The total number of devices including the computer is no more than 15.
• The total length of all the cables used is no more than 2 meter times the
number of devices connected together, up to a maximum of 20 meters.
N o t e
IEEE-488 states that you should exercise caution if your individual cable
lengths exceed 4 meters
Do not stack more than three connector blocks together on any GPIB
connector. Make sure that all connectors are fully seated and that the lock
screws are firmly finger tightened.