Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
76
INSTrument:COUPle[:TRIGger]?
This query returns the currently coupled output. Returns “ALL”, “NONE”, or
a list. If any output is not coupled, “NONE” is returned. If all of three
outputs are coupled, “ALL” is returned. If a list of outputs is coupled, the list
is returned.
Measurement Commands
MEASure:CURRent[:DC]? [{P6V | P25V | N25V}]
This command queries the current measured at the
output terminals
of the
power supply. The physical outputs of measurement are specified by the
output identifier. If any output identifier is not specified, the current of the
currently selected output is returned.
MEASure[:VOLTage][:DC]? [{P6V | P25V | N25V}]
This command queries the voltage measured at the
output terminals
of the
power supply. If any output identifier is not specified, the voltage of the
currently selected output is returned.
Содержание E3631-90002
Страница 9: ...8...
Страница 14: ...1 1 General Information...
Страница 25: ...24...
Страница 26: ...2 2 Initial Operation...
Страница 34: ...3 3 Front Panel Operation...
Страница 64: ...4 4 Remote Interface Reference...
Страница 113: ...112...
Страница 114: ...5 5 Error Messages...
Страница 124: ...6 6 Application Programs...
Страница 137: ...136...
Страница 138: ...7 7 Tutorial...
Страница 153: ...152...
Страница 154: ...8 8 Specifications...
Страница 160: ...Chapter 8 Specifications Supplemental Characteristics 159 8 Figure 8 1 Dimensions of Agilent E3631A Power Supply...
Страница 161: ...160...