Chapter 3 Front-Panel Operation
System-Related Operations
46
Display Control
For security reasons, you may want to turn off the front-panel display. From
the remote interface, you can display a 12-character message on the front
panel.
The display can be enabled / disabled from the remote interface only.
• When the display is turned off, outputs are not sent to the display and all
annunciators are disabled except the
ERROR
annunciator. Front-panel
operation is otherwise unaffected by turning off the display.
• The display state is stored in
volatile
memory; the display is always enabled
when power has been off, after a remote interface reset, or after returning to
local from remote.
• You can display a message on the front panel by sending a command from
the remote interface. The power supply can display up to 12 characters of
the message on the front panel; any additional characters are truncated.
Commas, periods, and semicolons share a display space with the preceding
character, and
are not
considered individual characters. When a message is
displayed, outputs are not sent to the display.
• Sending a message to the display from the remote interface overrides the
display state; this means that you can display a message even if the display is
turned off.
The display state is automatically turned on when you return to the local
(front panel) operation. Press the
key to return to the local state
from the remote interface
• Remote interface operation:
DISPlay {OFF|ON}
Disable / enable the display
DISPlay:TEXT
<quoted string>
Display the string
enclosed in quotes
DISPlay:TEXT:CLEar
Clear the displayed message
The following statement shows how to display a message on the front panel
from a Agilent Technologies controller.
"DISP:TEXT ’HELLO’"
Local
Содержание E3631-90002
Страница 9: ...8...
Страница 14: ...1 1 General Information...
Страница 25: ...24...
Страница 26: ...2 2 Initial Operation...
Страница 34: ...3 3 Front Panel Operation...
Страница 64: ...4 4 Remote Interface Reference...
Страница 113: ...112...
Страница 114: ...5 5 Error Messages...
Страница 124: ...6 6 Application Programs...
Страница 137: ...136...
Страница 138: ...7 7 Tutorial...
Страница 153: ...152...
Страница 154: ...8 8 Specifications...
Страница 160: ...Chapter 8 Specifications Supplemental Characteristics 159 8 Figure 8 1 Dimensions of Agilent E3631A Power Supply...
Страница 161: ...160...