![Agilent Technologies BioCel 1200 Скачать руководство пользователя страница 97](http://html1.mh-extra.com/html/agilent-technologies/biocel-1200/biocel-1200_user-manual_2867701097.webp)
91
Chapter 6: Maintenance and troubleshooting
BioCel User Guide
2.
In the
Robot Error
dialog box, select
Retry
.
Related topics
Resolving plate placing errors
About plate placing
errors
If the robot does not place a plate accurately on a device, the problem
could be with the:
❑
Teachpoint for either that device or for the previously scheduled
device, including approach height.
In this case, see for more information about fine-tuning teachpoints.
❑
Device, because it was moved or reconfigured by someone else who
did not re-teach the teachpoints.
❑
Plate.
In this case, replace the plate with one that is not deformed.
Relate information
For information about...
See...
Opening a door during a run
“Pausing and stopping a run” on
page 60
Safety interlocks and emergency
stops
“Comparing interlock and emergency
circuits” on page 62
For information about...
See...
Fine-tuning teachpoints
Содержание BioCel 1200
Страница 1: ...Agilent Technologies BioCel System User Guide...
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Страница 18: ...Chapter 1 Introduction BioCel User Guide 12...
Страница 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Страница 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Страница 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Страница 104: ...Index BioCel User Guide 98...
Страница 105: ......
Страница 106: ...Agilent Technologies User Guide G5500 90002...