![Agilent Technologies BioCel 1200 Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/agilent-technologies/biocel-1200/biocel-1200_user-manual_2867701080.webp)
Chapter 5: Diagnostics
BioCel User Guide
74
3.
In the
Position Control
area, click one of the following buttons:
The readout displays the current coordinates of the robot in relation
to the home position.
Moving the gripper
You can open and close the robot’s gripper in a single movement.
To open or close the robot’s gripper:
1.
In the
Position Control
area, click
Open gripper
or
Close gripper
.
Setting
Comments
Gripper offset
The height above the teachpoint, which is
typically 2–3 mm.
You will rarely need to change this value from
the values automatically entered for the plate
type. However, you can type a value into the
Gripper offset text box if you need to.
Be aware however, that this is a temporary
change. To change it permanently, you need to
make the change in the labware editor.
Jog increment
The distance or angle to move the robot at each
jog.
Button
Moves the robot...
Down
Down, decreasing the
z
-axis value
Up
Up, increasing the
z
-axis value
In
In towards the center of the robot’s
radius, decreasing the
r
-axis value
Out
Out from the center of the robot’s
radius, increasing the
r-
axis value
CW
Clockwise, decreasing the
t
-axis value
CCW
Counterclockwise, increasing the
t-
axis value
Содержание BioCel 1200
Страница 1: ...Agilent Technologies BioCel System User Guide...
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Страница 18: ...Chapter 1 Introduction BioCel User Guide 12...
Страница 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Страница 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Страница 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Страница 104: ...Index BioCel User Guide 98...
Страница 105: ......
Страница 106: ...Agilent Technologies User Guide G5500 90002...