Chapter 3: Preparing for a run
BioCel User Guide
Performing pre-run checks
About this topic
After turning on the BioCel, creating a protocol, and preparing the
devices you would typically perform a number of pre-run checks. This
topic describes the checks to perform.
Procedure
To check that the BioCel is ready for a run:
1.
For each VPrep, make sure that any reservoirs and Micro Wash trays
are on the correct shelves.
2.
Make sure that there are no stray plates or lids in robot-accessible
places, which include:
Platepads
VPrep shelves
VStack plate stages
VSpin buckets
VCode plate stages
PlateLoc plate stages
Plate hotel shelves
PlateHubs
Lid hotel bays
Lid removal devices
Third-party devices
You can also set the robot to check all accessible positions at the
beginning of a run.
!! DAMAGE HAZARD !!
Leaving a stray plate in the system
might cause the robot, when holding a plate, to crash into the
stray plate. A typical robot crash will damage the plates, spill
samples and stop the robot. It might also damage the robot
gripper pads.
Workflow this procedure belongs
to
“Workflow for preparing the BioCel for
a Run” on page 42
Preparing devices for a run
“Preparing devices and accessories” on
page 46
For information about...
See...
Содержание BioCel 1200
Страница 1: ...Agilent Technologies BioCel System User Guide...
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Страница 18: ...Chapter 1 Introduction BioCel User Guide 12...
Страница 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Страница 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Страница 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Страница 104: ...Index BioCel User Guide 98...
Страница 105: ......
Страница 106: ...Agilent Technologies User Guide G5500 90002...