![Agilent Technologies BioCel 1200 Скачать руководство пользователя страница 67](http://html1.mh-extra.com/html/agilent-technologies/biocel-1200/biocel-1200_user-manual_2867701067.webp)
61
Chapter 4: Performing a run
BioCel User Guide
❑
Cuts power to the VPrep motors that move the head, and activates
the brake on the head to prevent it from falling
❑
Stops the air supply to the instruments
Opening an
enclosure door
during a run
Provided the interlock circuit has not been overridden, you can open an
enclosure door at any time during a run to instantly stop BioCel motion.
The interlock light on the power panel will turn off when this happens
and generates a robot error, after which you can either continue or abort
the run or enter diagnostics.
When you close the door, an error dialog box opens.
If a VPrep was in motion when the door was opened, a VPrep Error
dialog box might also be displayed.
!! IMPORTANT !!
Do not use the door interlocks to routinely stop
a protocol. The door interlocks are designed to be a safety backup
system. Instead, use the pause button in VWorks.
Related topics
00005
For more information about...
See...
Working with diagnostics software
Recovering from an emergency
stop
“Emergency stop error recovery” on
page 87.
Recovering from an interlock error
Содержание BioCel 1200
Страница 1: ...Agilent Technologies BioCel System User Guide...
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Страница 18: ...Chapter 1 Introduction BioCel User Guide 12...
Страница 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Страница 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Страница 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Страница 104: ...Index BioCel User Guide 98...
Страница 105: ......
Страница 106: ...Agilent Technologies User Guide G5500 90002...