81
Chapter 5: Diagnostics
BioCel User Guide
The next diagram shows a side view of the correct teachpoint to
illustrate the correct
z
-teachpoint.
Incorrect teachpoint example
In the next diagram, the teachpoint is incorrect. The back of the robot
gripper stops before aligning with the back of the teachplate calibration
tool, and the teachplate is held too far back in the gripper. The result is
the plate is delivered too far forward and thus does not sit correctly on
the platepad.
00259
Correct Teachpoint
00259
Optimal teachpoint
Back of plate
00261
Side Teachpoint
00261