Chapter 3: Preparing for a run
BioCel User Guide
3.
If the radius width of the label backing on the take-up roll is more
than 1 inch (2.5 cm) wide, remove the label backing from the take-
up roll, cut it, and re-attach the end.
Too much label backing on the take-up roll can interfere with
operation of the VCode.
4.
Use VCode Diagnostics to perform a plate-labeling test.
VSpin set up
For more detailed information about operating the VSpin manually, see
the
VSpin User Guide
.
The VSpin can spin one or two sample plates at the same time. Spinning
one sample plate at a time always requires a counterweight plate. The
counterweight plate can be added manually or robotically.
To set up a VSpin:
1.
Do one of the following:
If you are using a manually loaded centrifuge counterweight,
place the counterweight into bucket 2, using VSpin Diagnostics
to open and close the VSpin door.
If you are using robot-loaded counterweights, make sure you
have correctly set up counterweights in VWorks, placed the
counterweights in the correct plate positions, and checked to
make sure the buckets are empty.
If you are spinning two protocol plates and not using a
counterweight, make sure you have set up the plate loading
properly in VWorks and that both buckets are empty.
!! DAMAGE HAZARD !!
To avoid a robot crash, always place the
counterweight into bucket 2 and leave the door open with bucket 1
at the front. The sample plate will always be moved to bucket 1 by
the robot.
Related topics
For more information about...
See...
Setting up a counterweight
“Setting up a VSpin counterweight
plate” on page 51
Workflow this procedure belongs
to
“Workflow for preparing the BioCel for
a Run” on page 42
Testing an auxiliary bar code
reader
“Aligning and testing auxiliary bar code
readers” on page 53
Placing and removing hotel plates
“Placing and removing plate hotel
labware” on page 50
Setting up a PlateHub
Содержание BioCel 1200
Страница 1: ...Agilent Technologies BioCel System User Guide...
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division...
Страница 18: ...Chapter 1 Introduction BioCel User Guide 12...
Страница 64: ...Chapter 3 Preparing for a run BioCel User Guide 58...
Страница 74: ...Chapter 4 Performing a run BioCel User Guide 68...
Страница 100: ...Chapter 6 Maintenance and troubleshooting BioCel User Guide 94...
Страница 104: ...Index BioCel User Guide 98...
Страница 105: ......
Страница 106: ...Agilent Technologies User Guide G5500 90002...