119
3
N9020A
Midrange Signal Analyzer, MXA (cont.)
Signal Analysis
English URL
www.agilent.com/find/products
Second Harmonics Distortion (SHI)
Mixer level
Distortion
SHI
10 MHz to 1.8 GHz
–15 dBm
–60 dBc
+45 dBm
1.8 to 7.0 GHz
–15 dBm
–80 dBc
+65 dBm
7.0 to 11.0 GHz
–15 dBm
–70 dBc
+55 dBm
11.0 to 13.25 GHz
–15 dBm
–65 dBc
+50 dBm
Preamp On (Option P03, P08, P13, P26)
Preamp level Distortion
SHI
10 MHz to 1.8 GHz
–45 dBm
–78 dBc nominal
+33 dBm nominal
1.8 to 13.25 GHz
–50 dBm
–60 dBc nominal
+10 dBm nominal
Third-order Intermodulation Distortion (TOI)
(two –30 dBm tones at input mixer with tone separation >5 times
IF prefilter bandwidth, 20 to 30°C, see Specifications Guide for
IF prefilter bandwidths)
Distortion
TOI
Typical
10 to 100 MHz
–84 dBc
+12 dBm
+17 dBm
100 to 400 MHz
–88 dBc
+14 dBm
+18 dBm
400 MHz to 1.7 GHz –90 dBc
+15 dBm
+19 dBm
1.7 to 3.6 GHz
–92 dBc
+16 dBm
+19 dBm
3.6 to 8.4 GHz
–90 dBc
+15 dBm
+18 dBm
8.4 to 13.6 GHz
–90 dBc
+15 dBm
+18 dBm
13.6 to 26.5 GHz
–80 dBc
+10 dBm
+14 dBm
Preamp on (Option P03, P08, P13, P26)
(two –45 dBm tones at preamp input)
10 to 500 MHz
+4 dBm nominal
500 MHz to 3.6 GHz
+5 dBm nominal
3.6 to 26.5 GHz
–15 dBm nominal
Phase Noise
Noise sidebands (20 to 30°C, CF = 1 GHz)
Offset
Specification
Typical
100 Hz
–84 dBc/Hz
–88 dBc/Hz
1 kHz
–100 dBc/Hz nominal
10 kHz
–103 dBc/Hz
–106 dBc/Hz
100 kHz
–115 dBc/Hz
–117 dBc/Hz
1 MHz
–133 dBc/Hz
–137 dBc/Hz
10 MHz
–148 dBc/Hz nominal
Power Suite Measurement Specifications
Channel Power
Amplitude accuracy, W-CDMA or IS-95 (20 to 30°C, attenuation = 10 dB):
±0.80 dB (±0.30 dB 95
th
percentile)
Occupied Bandwidth (OBW)
Frequency accuracy
±[span/1000] nominal
Adjacent Channel Power (ACP)
Accuracy, W-CDMA (ACLR)
(at specific mixer levels
and ACLR ranges)
Adjacent
Alternate
MS
±0.14 dB
±0.21 dB
BTS
±0.49 dB
±0.44 dB
Dynamic range (typical)
Without noise correction
–73 dB
–79 dB
With noise correction
–78 dB
–82 dB
Offset channel pairs measured 1 to 6
ACP speed (fast method). Data measurement and transfer time:
14 ms nominal (
= 0.2 dB)
ACPR dynamic range, W-CDMA (5 MHz offset, RRC weighted, 3.84 MHz
noise bandwidth)
Two carriers
–70 dB nominal
Four carriers
–64 dB nominal
With noise correction
–72 dB nominal
ACPR accuracy (two carriers, 5 MHz offset, –48 dBc ACPR):
±0.42 dB nominal
Multiple number of carriers measured: Up to 12
Power Statistics CCDF
Histogram resolution
0.01 dB
Spurious Emission
W-CDMA (1 to 3.6 GHz)
Table driven spurious signals; search across regions.
Dynamic range
95.3 dB (100.3 dB typical)
Absolute sensitivity
–84.4 dBm (–89.4 dBm typical)
Spectrum Emission Mask (SEM)
cdma2000 (750 kHz offset)
Relative dynamic range
(30 kHz RBW)
78.9 dB (85.0 dB typical)
Absolute sensitivity
–99.7 dBm (–104.7 dBm typical)
Relative accuracy
±0.11 dB
3GPP W-CDMA (2.515 MHz offset)
Relative dynamic range
(30 kHz RBW)
81.9 dB (88.2 dB typical)
Absolute sensitivity
–99.7 dBm (–104.7 dBm typical)
Relative accuracy
±0.12 dB
General Specifications
Temperature Range
Operating
5 to +50°C
Storage
–40 to +65°C
EMC
Complies with European EMC Directive 89/336/EEC,
amended by 93/68/EEC
IEC/EN 61326
CISPR Pub 11 Group 1, class A
AS/NZS CISPR 11:2002
ICES/NMB-001
Safety
Complies with European Low Voltage Directive 73/23/EEC,
amended by 93/68/EEC
IEC/EN 61010-1
Canada:CSA C22.2 No.61010-1
USA:UL 61010-1
Audio Noise
Acoustic noise emission
Geraeuschemission
LpA <70 dB
LpA <70 dB
Operator position
Am Arbeitsplatz
Normal position
Normaler Betrieb
Per ISO 7779
Nach DIN 45635 t.19
Environmental Stress
Samples of this product have been type tested in accordance with the
Agilent Environmental Test Manual and verified to be robust against the
environmental stresses of storage, transportation and end-use; those
stresses include but are not limited to temperature, humidity, shock,
vibration, altitude and power line conditions. Test methods are aligned
with IEC 60068-2 and levels are similar to MIL-PRF-2800F Class 3.
Power Requirements
Voltage and frequency (nominal)
100/120 V, 50/60 Hz
220/240 V, 50/60 Hz
Power Consumption
On
< 260 watts
Standby
< 20 watts
Data Storage
Internal
40 GB nominal
External
Supports USB 2.0
compatible memory devices
Weight (without options)
Net
16 kg (35 lbs) nominal
Shipping
28 kg (62 lbs) nominal
Dimensions
Height
177 mm (7.0 in)
Width
426 mm (16.8 in)
Length
368 mm (14.5 in)
Warranty
The MXA signal analyzer is supplied with a one-year warranty.
Calibration Cycle
The recommended calibration cycle is one year: Calibration services are
available through Agilent service centers.
Содержание All
Страница 1: ......