![Agilent Technologies All Скачать руководство пользователя страница 175](http://html.mh-extra.com/html/agilent-technologies/all/all_catalog_2868003175.webp)
222
3
English URL
www.agilent.com/find/products
English URL
www.agilent.com/find/products
34401A
English URL
www.agilent.com/find/products
34401A
Low-Cost 6
1
⁄
2
Digit Multimeter
Digital Multimeters
34401A Digital Multimeter
The 34401A digital multimeter establishes a new price/performance
standard by offering such features as 6
1
⁄
2
digits of resolution, 1000
readings per second, and 15 ppm basic dc accuracy at a surprisingly
affordable price. The 34401A has been designed for superior perfor-
mance while providing the flexibility to meet both your present and
future needs.
Great Bench Performance
The clear, logical front panel of the 34401A allows you to easily
select all primary measurement functions. Traditional “bench”
functions, such as continuity and diode test, are included. Math
functions, such as NULL, dB, dBm, limit test, and min/max/avg are
easily selected. A simple menu scheme gives you access to powerful
advanced features, such as the ability to store up to 512 readings in
internal memory. Measurement results are displayed on a bright,
high-visibility readout. A rugged case ensures survival even under
the toughest conditions, and the optional accessory pouch makes it
easy to pack up and go with the 34401A.
Superior Performance in Your System
The 34401A can take up to 1000 readings per second, including
GPIB bus transfer in ASCII format. Both GPIB and RS-232 are
standard, letting you select the interface that best meets your needs.
The 34401A responds to three different command languages. It
accepts SCPI commands (Standard Commands for Programmable
Instruments), which ensures present and future compatibility.
Drivers are also available for both National Instruments Labview
and Agilent’s VEE software.
IntuiLink Software, included with your 34401A, allows you to
transfer your measurement data and images into Microsoft Excel or
Microsoft Word with little or no programming. You can specify the
meter setup and take a single reading or log data to the Excel
spreadsheet in specific time intervals. Programmers can use the
ActiveX components to control the DMM using SCPI commands. To
find out more about IntuiLink, visit
www.agilent.com/find/intuilink
• 12 measurement functions
• 1000 V maximum input
• 15 ppm basic dcV accuracy (24-hour)
• 1000 readings per second direct to GPIB
• GPIB and RS-232 standard
• 512-reading memory
• SCPI commands standard
• IntuiLink Connectivity Software included
Abbreviated Technical Specifications
DC Voltage
Input Characteristics
Range
Maximum
Resolution in Digits
Input Resistance
Reading
6
1
⁄
2
5
1
⁄
2
4
1
⁄
2
(6
1
⁄
2
digits)
100 mV
120.0000
100 nV 1 µV
10 µV
10 M
Ω
or >10 G
Ω
1 V
1.200000
1 µV
10 µV
100 µV
10 M
Ω
or >10 G
Ω
10 V
12.00000
10 µV
100 µV 1 mV
10 M
Ω
or >10 G
Ω
100 V
120.0000
100 µV 1 mV
10 mV
10 M
Ω
1000 V
1050.000
1 mV
10 mV 100 mV
10 M
Ω
Input Protection:
>1000 V on all ranges
Measurement Accuracy:
±(% of r % of range)
Range
24-Hour
90-Day
1-Year
23°C ± 1°C
23°C ± 5°C
23°C ± 5°C
100 mV
0.0030 + 0.0030
0.0040 + 0.0035
0.0050 + 0.0035
1 V
0.0020 + 0.0006
0.0030 + 0.0007
0.0040 + 0.0007
10 V
0.0015 + 0.0004
0.0020 + 0.0005
0.0035 + 0.0005
100 V
0.0020 + 0.0006
0.0035 + 0.0006
0.0045 + 0.0006
1000 V
0.0020 + 0.0006
0.0035 + 0.0010
0.0045 + 0.0010
Noise Rejection:
(50 or 60 Hz, 1 k
Ω
unbalance in LO lead)
DC CMRR:
140 dB
AC CMRR:
70 dB
Normal Mode Rejection
(60 Hz/50 Hz) ± 0.1%
:
100 PLC (1.67 s/2.00 s): 60 dB
10 PLC (167 ms/200 ms): 60 dB
1 PLC (16.7 ms/20.0 ms): 60 dB
<1 PLC (3 ms or 800
µ
s): 0 dB
Maximum Reading Rate:
(readings/s)
Power Line Frequency
Resolution in Digits
6
1
⁄
2
5
1
⁄
2
4
1
⁄
2
60 Hz
6
300
1000
50 Hz
5
300
1000