76
86106B
86116A
86116B
86116C
English URL
www.agilent.com/find/products
3
86100C Infiniium DCA-J (cont.)
Optical/Transmission
Module Specifications: 86106B, 86116A/B/C
High Bandwidth, Single-mode Optical/
86106B
86116A
1
86116B
1
86116C
1
Electrical Modules
Optical Channel Specifications
Optical Channel Unfiltered Bandwidth
28 GHz
53 GHz
65 GHz (best pulse fidelity)
65 GHz
Wavelength Range
1000 to 1600 nm
1480 to 1620 nm
1480 to 1620 nm
Calibrated Wavelengths
1310/1550 nm
1550 nm
1550 nm
Optical Sensitivity
3
–7 dBm
–3 dBm
Transition Time
(10% to 90%, calculated
18 ps
9.0 ps (FWHM)
2
7.4 ps (FWHM)
2
7.4 ps (FWHM)
2
from TR = 0.48/BW optical)
RMS Noise
Characteristic
13 µW (Filtered)
60 µW (50 GHz)
50 µW (55 GHz)
36 µW (39.8, 43.0 Gb/s filters)
23 µW (Unfiltered)
190 µW (53 GHz)
140 µW (65 GHz)
125 µW (65 GHz)
Maximum
15 µW (Filtered)
90 µW (50 GHz)
85 µW (55 GHz)
68 µW (39.8, 43.0 Gb/s filters)
30 µW (Unfiltered)
260 µW (53 GHz)
250 µW (65 GHz)
200 µW (65 GHz)
Scale Factor
Minimum
20 µW/division
200 µW/division
200 µW/division
200 µW/division
Maximum
500 µW/division
2.5 mW/division
5 mW/division
5 mW/division
CW Accuracy
(single marker, referenced
±50 µW ± 4% of
±150 µW ± 4% of
±150 µW ± 4% of
±150 µW ± 4% of
to average power monitor)
(reading-channel offset)
(reading-channel offset)
(reading-channel offset)
(reading-channel offset)
CW Offset Range
(referenced two
+1 mW to –3 mW
+5 mW to –15mW
+8 to –12 mW
+8 to –12 mW
divisions from screen bottom)
Average Power Monitor
–27 dBm to +3 dBm
–23 dBm to +9 dBm
–23 dBm to +9 dBm
–23 dBm to +9 dBm
(specified operating range)
Factory Calibrated Accuracy
±5% ± 100 nW ± connector uncertainty, 20°C to 30°C
User Calibrated Accuracy
±2% ± 100 nW ± power meter uncertainty, <5°C change
Maximum Input Power
Maximum non-destruct average
2 mW (+3 dBm)
10 mW (+10 dBm)
10 mW (+10 dBm)
10 mW (+10 dBm)
Maximum non-destruct peak
10 mW (+10 dBm)
50 mW (+17 dBm)
50 mW (+17 dBm)
50 mW (+17 dBm)
Fiber Input
9/125 µm, user selectable connector
Input Return Loss
30 dB
30 dB
20 dB
20 dB
(HMS-10 connector fully filled fiber)
Electrical Channel Specifications
Electrical Channel Bandwidth
18 and 40 GHz
43 and 63 GHz
80, 55 and 30 GHz
80, 55 and 30 GHz
Transition Time
(10% to 90%,
19.5 ps (18 GHz)
8.1 ps (43 GHz)
6.4 ps (55 GHz)
6.4 ps (55 GHz)
calculated from TR = 0.35/BW)
9 ps (40 GHz)
5.6 ps (63 GHz)
4.4 ps (80 GHz)
4.4 ps (80 GHz)
RMS Noise
Characteristic
0.25 mV (18 GHz)
0.6 mV (43 GHz)
0.6 mV (55 GHz)
0.5 mV (30 GHz)
0.5 mV (40 GHz)
1.7 mV (63 GHz)
1.1 mV (80 GHz)
1.1 mV (80 GHz)
Maximum
0.5 m V (18 GHz)
0.9 mV (43 GHz)
1.1 mV (55 GHz)
0.8 mV (30 GHz)
1.0 mV (40 GHz)
2.5 mV (63 GHz)
2.2 mV (80 GHz)
2.2 mV (80 GHz)
Scale Factor
Minimum
1 mV/division
2 mV/division
2 mV/division
2 mV/division
Maximum
100 mV/division
100 mV/division
100 mV/division
100 mV/division
DC Accuracy
(single marker)
±0.4% of full scale
±0.8% of full scale
±0.4% of full scale
±0.4% of full scale
±2 mV ± 1.5% of (reading-
±2 mV ± 1.5% of (reading-
±3 mV ± 2% of (reading-
±3 mV ± 2% of (reading-
channel offset), 18 GHz
channel offset), 43 GHz
channel offset), ± 2% of
channel offset), ± 2% of
±0.4% of full scale
±2.5% of full scale
offset (all bandwidths)
offset (all bandwidths)
±2 mV ± 3% of (reading-
±2 mV ± 2% of (reading-
channel offset), 40 GHz
channel offset), 63 GHz
DC Offset Range
(referenced to
±500 mV
±500 mV
±500 mV
±500 mV
center of screen)
Input Dynamic Range
(relative to
±400 mV
±400 mV
±400 mV
±400 mV
channel offset)
Maximum Input Signal
±2 V (+16 dBm)
±2 V (+16 dBm)
±2 V (+16 dBm)
±2 V (+16 dBm)
Nominal Impedance
50
Ω
50
Ω
50
Ω
50
Ω
Reflections
(for 20 ps rise time)
5%
5%
10% (DC to 70 GHz)
10% (DC to 70 GHz)
20% (70 to 100 GHz)
20% (70 to 100 GHz)
Electrical Input
2.4 mm (male)
1.85 mm (male)
1.85 mm (male)
1.85 mm (male)
1
86116A and 86116B require the 86100 software revision A.3.0 or above. 86116C requires an 86100C mainframe and software revision 7.0.
2
FWHM (Full Width Half Max) as measured from optical pulse with 700 fs FWHM, 5 MHz repetition rate and 10 mW peak power.
3
Smallest average optical power required for mask test. Values represent typical sensitivity of NRZ eye diagrams. Assumes mask test with compliance filter switched in.
Содержание All
Страница 1: ......