248
3
English URL
www.agilent.com/find/products
53150A
53151A
53152A
CW Microwave Frequency Counters (cont.)
Electronic Counters
53150 Series Abridged Measurement Specifications
and Characteristics
All measurement specifications are over the full signal ranges of
channels 1 and 2. For full specifications or request a Product
Overview for the 53150 series microwave counters, please visit our
web site
www.agilent.com/find/frequencycounters
Input Characteristics
Input 1 (1 M
Ω
)
Input 2 (50
Ω
)
Frequency Range
53150A
10 Hz to 125 MHz
0.05 – 20 GHz
53151A
10 Hz to 125 MHz
0.05 – 26.5 GHz
53152A
10 Hz to 125 MHz
0.05 – 46 GHz
Sensitivity
53150/51
53152
<30 Hz
40 mV
rms
—
—
to 125 MHz
25 mV
rms
—
—
<300 MHz
—
–20 dBm
–20 dBm
to 12.4 GHz
—
–33 dBm
–33 dBm
to 18 GHz
—
–33 dBm
–30 dBm
to 20 GHz
—
–29 dBm
–27 dBm
to 26.5 GHz
—
–25 dBm (151)
–27 dBm
to 40 GHz
—
—
–23 dBm
to 46 GHz
—
—
–17 dBm
Maximum Input
2 V
rms
+5 dBm, <2 GHz
+13 dBm, >2 GHz
Damage Level
5 V
rms
to 120 Vp
+27 dBm
Coupling
AC
AC
Channel 1
Resolution:
Selectable 1 Hz to 1 MHz
Connector:
BNC female
Low Pass Filter:
50 kHz, selectable
Channel 2
Resolution:
Selectable 1 Hz to 1 MHz
Acquisition Time:
100 ms to 140 ms
Gate Time:
1/Resolution
FM Tolerance:
20 MHz p-p max. @10 MHz rate to 26.5 GHz;
12 MHz p-p max. @10 MHz rate to above 26.5 GHz
AM Tolerance:
Any depth/rate within dynamic range of input
Amplitude Discrimination
(above 250 MHz)
:
20 dB typ. for greater than
75 MHz separation; 10 dB typ. for less than 75 MHz separation
Connector:
3.5 mm SMA compatible (53150A/53151A);
2.92 mm removable (53152A)
Power Measurement
Range:
Counter sensitivity to +7 dBm
Units:
dBm or milliwatts/microwatts
Resolution:
0.01 dB
Accuracy*
(0 to –20 dBm)
:
53150/51
53152
<12.4 GHz
±1.5 dB
±1.0 dB
to 20 GHz
±1.5 dB
±1.5 dB
to 26.5 GHz
±2.0 dB (151)
±1.5 dB
to 46 GHz
—
±2.0 dB
* At channel two input connector
General
Display:
Backlit LCD
Sample Rate:
User-selectable Fast, Medium, Slow or Hold
Programming:
GPIB and RS-232C, SCPI compatible
Math Functions
• Offset: Last reading and/or entered offset to reading for either
power or frequency
• Averaging: 1 to 99 measurement running average
• Power Correction (Cable Loss Compensation): Offsets
power reading via linear interpolation of up to 10 user-entered
frequency attenuations
Save and Recall:
Up to 9 complete instrument setups may be saved
and later recalled
Sleep Mode
(battery option only)
:
Automatically activated if no input
is present for 5 minutes
Power Requirements
• 75 VA max. (25 W typ.)
• 90 – 132 VAC; 50, 60, 400 Hz
• 216 – 264 VAC; 50, 60 Hz
• 11 – 18 VDC: 2 A max.: battery option only
Battery
(option)
:
• Type: VHS camcorder, sealed lead acid (2 each)
• Charge Time: 8 hours in unit
• Capacity: 2.5 hours minimum at 25°C
Timebase:
TCX0 (std)
Option 001
Aging Rate
1 x 10
–7
/mo
5 x 10
–10
/day
Short Term (1 sec. avg.)
1 x 10
–9
2 x 10
–10
Temperature (0° – 55°C)
<1 x 10
–6
1 x 10
–8
Accessories Furnished:
Power cord and operating, programming,
and service manuals
Size:
88.5 mm H x 213 mm W x 300 mm D (3.54 in x 8.52 in x 12 in)
Weight:
4 kg (8.8 lb) without battery option; 6.4 kg (14.08 lb) with
battery option
Ordering Information
53150A
20 GHz Microwave Counter
53151A
26.5 GHz Microwave Counter
53152A
46 GHz Microwave Counter
5315xA-001
Oven Timebase
5315xA-002
Battery/DC Input
5315xA-007
Soft Carrying Case
5315xA-A6J
ANSI Z540 Compliant Calibration
5315xA-1CM
Rack Mounting Kit
Содержание All
Страница 1: ......