Vacuum Gauge Controller AGC-100
80
7 Troubleshooting
The error is shown in the dot matrix and the error relay opens (
→
20).
Possible cause and remedy/acknowledgement
SENSOR ERROR
I
Interruption or instability in sensor line or connector
(Sensor error).
Acknowledge with the
key.
WATCHDOG ERROR
I
The AGC-100 has been turned on too fast after power
off.
Acknowledge with the
key.
If the watchdog is set to
Auto
, the AGC-100 ack-
nowledges the message automatically after 2 s
(
→
48).
The watchdog has tripped because of a severe electric
disturbance or an operating system error.
Acknowledge with the
key.
If the watchdog is set to
WATCHDOG AUTO
,
the AGC-100 acknowledges the message auto-
matically after 2 s (
→
48).
DATA CORRUPTED
I
Parameter memory error (EEPROM).
Acknowledge with the
key.
If the problem persists after the message has been acknowledged several
times and/or the gauge has been exchanged, please contact your nearest
Agilent Technologies service center.
Signalization of errors
Error messages
Technical support