Vacuum Gauge Controller AGC-100
77
Transmit:
TMP
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: aa
<CR><LF>
Description
aa
Temperature (±2 °C) [°C]
Transmit:
TRS
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit:
<ENQ> Starts the test (repeats each character, test is interrupted
with
<CTRL>
C).
Transmit:
WDT
[,a] <CR>[<LF>]
Description
a
Watchdog control, a =
0 –> Manual error acknowledgement
1 –> Automatic error acknowledgement
1)
(default)
1)
If the watchdog has responded, the error is automatically acknowl-
edged and cancelled after 2 s.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a
<CR><LF>
Description
a Watchdog
control
Transmit:
AYT
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,b,c,d,e
<CR><LF>
Description
a
Type of the unit, e.g. AGC-100
b
Model No. of the unit, e.g. AGC100
c
Serial No. of the unit, e.g. LI1605F001
d
Firmware version of the unit, e.g. 1.00
e
Hardware version of the unit, e.g. 1.0
TMP
- Inner Temperature of
the Unit
TRS
- Serial Interface test
WDT
- Watchdog control
5.12 Further
AYT
- Are you there?