
Vacuum Gauge Controller AGC-100
61
Transmit:
SPx
[,a,x.xxxxEsxx,y.yyyyEsyy] <CR>[<LF>]
Description
x
Switching function, x =
1 –> Switching function 1
2 –> Switching function 2
a
Switching function assignment, a =
0 –> Turned off
1 –> Turned on
2 –> Measurement channel 1
x.xxxxEsxx Lower
threshold
1)
[in current pressure unit]
(default = depending on gauge) (s = sign)
y.yyyyEsyy Upper
threshold
1)
[in current pressure unit]
(default = depending on gauge) (s = sign)
1)
Values can be entered in any format.
They are internally converted into the floating point format.
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a,x.xxxxEsxx,y.yyyyEsyy
<CR><LF>
Description
a
Switching function assignment
x.xxxxEsxx
Lower threshold [in current pressure unit]
(s = sign)
y.yyyyEsyy
Upper threshold [in current pressure unit]
(s = sign)
Transmit:
COR
[,a.aaa] <CR>[<LF>]
Description
a.aaa
Calibration factor, 0.100 … 10.000 (default = 1.000)
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: a.aaa,b.bbb,c.ccc
<CR><LF>
Description
a.aaa
Calibration factor
SP1, SP2
- Switching function 1
and 2
5.6 Gauge Parameters
COR
- Calibration factor
Содержание AGC-100
Страница 2: ...Va AGC acuu C 100 um G Gaug ge Co Cop Ag ontro pyright 2016 gilent Inc oller ...
Страница 3: ......
Страница 24: ...Vacuum Gauge Controller AGC 100 24 Bargraph FSR ...
Страница 92: ...Vacuum Gauge Controller AGC 100 92 Notes ...
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...t qna96e1 ...