Vacuum Gauge Controller AGC-100
35
The measurement value filter permits a better evaluation of unstable or disturbed
measuring signals.
Value
FILTER
I
FILTER OFF
I
No measurement value filter
FILTER FAST
I
Fast:
The AGC-100 responds quickly to fluctua-
tions in the measurement value. As a result,
it will respond faster to interference in the
measured values.
FILTER NORMAL
I
Normal (factory setting):
Good relationship between response and
sensitivity of the display and the switching
function to changes in the measured values.
FILTER SLOW
I
Slow:
The AGC-100 does not respond to small
changes in measured values. As a result, it
will respond more slowly to changes in the
measured values.
Pressure p
Time t
Pressure p
Time t
Pressure p
Time t
Measurement value filter