Vacuum Gauge Controller AGC-100
75
Transmit:
TDI
[,a] <CR>[<LF>]
Description
a
Display test, a =
0 –> Stops the test - display according to current operating
mode (default)
1 –> Starts the test - all segments on
Receive: <ACK><CR><LF>
Transmit: <ENQ>
Receive: x
<CR><LF>
Description
x
Display test status
Test of the parameter memory.
Transmit:
TEE
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit:
<ENQ>
Starts the test (duration <1 s)
Do not keep repeating the test (EEPROM life).
Receive: aaaa
<CR><LF>
Description
aaaa Error
word
Test of the program memory.
Transmit:
TEP
<CR>[<LF>]
Receive: <ACK><CR><LF>
Transmit:
<ENQ>
Starts the test (very brief)
Receive: aaaa,bbbbbbbb
<CR><LF>
Description
aaaa Error
word
bbbbbbbbbb Check sum (hex)
Caution
Caution: The relays switch irrespective of the pressure.
Starting a test program may cause unwanted effects in connected
control systems.
Disconnect the sensor cable and control system lines to ensure that
no control commands or messages are triggered by mistake.
TDI
- Display test
TEE
- EEPROM test
TEP
- FLASH test
TIO
- I/O test