Vacuum Gauge Controller AGC-100
55
Messages are transmitted to the AGC-100 as ASCII strings in the form of mne-
monic operating codes and parameters. All mnemonics comprise three ASCII
characters.
Spaces are ignored. <ETX> (CTRL C) clears the input buffer in the AGC-100.
HOST AGC-100
Explanation
Mnemonics [and parameters] –––––––––>
<CR>[<LF>] ––––––––––––––––––––––>
Receives message with "end of
message"
<––––––––––––––––– <ACK><CR><LF>
Positive acknowledgment of a re-
ceived message
When requested with a mnemonic instruction, the AGC-100 transmits the measure-
ment data or parameters as ASCII strings to the HOST.
<ENQ> (CTRL E) must be transmitted to request the transmission of an ASCII
string. Additional strings, according to the last selected mnemonic, are read out by
repetitive transmission of <ENQ>.
If <ENQ> is received without a valid request, the ERROR word is transmitted.
HOST AGC-100
Explanation
Mnemonics [and parameters] –––––––––>
<CR>[<LF>] ––––––––––––––––––––––>
Receives message with "end of
message"
<–––––––––––––––––- <ACK><CR><LF> Positive acknowledgment of a re-
ceived message
<ENQ> -––––––––––––––––––––––––>
Requests to transmit data
<––––––––––––––– Measurement values
or
parameters
<––––––––––––––––––––––– <CR><LF>
Transmits data with "end of mes-
sage"
:
:
<ENQ> –––––––––––––––––––––––––>
Requests to transmit data
<––––––––––––––– Measurement values
or
parameters
<––––––––––––––––––––––– <CR><LF>
Transmits data with "end of mes-
sage"
The strings received are verified in the AGC-100. If an error is detected, a negative
acknowledgment <NAK> is output.
HOST AGC-100
Explanation
Mnemonics [and parameters] –––––––––>
<CR>[<LF>] -–––––––––––––––––––––>
Receives message with "end of
message"
***** Transmission or programming error *****
<–––– <NAK><CR><LF>
Negative acknowledgment of a re-
ceived message
Mnemonics [and parameters] ––––––––->
<CR>[<LF>] –––––––––––––––––––––>
Receives message with "end of
message"
<–––––––––––––––– <ACK><CR><LF>
Positive acknowledgment of a re-
ceived message
5.2 Communication
Protocol
Transmission format
Transmission protocol
Reception format
Reception protocol
Error processing
Error recognition protocol
Содержание AGC-100
Страница 2: ...Va AGC acuu C 100 um G Gaug ge Co Cop Ag ontro pyright 2016 gilent Inc oller ...
Страница 3: ......
Страница 24: ...Vacuum Gauge Controller AGC 100 24 Bargraph FSR ...
Страница 92: ...Vacuum Gauge Controller AGC 100 92 Notes ...
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...t qna96e1 ...