Vacuum Gauge Controller AGC-100
51
Decimal separator for measurement values in the measurement data file.
Wert
DEC-SEPARATOR
I
DEC-SEPARATOR ,
I
Decimal comma
DEC-SEPARATOR .
I
Decimal point
Value
FILENAME
I
Name of the measurement data file, max.
7 digits
e.g.
FILENAME DATALOG
I
File ending: CSV
After entering the 7
th
digit the display stops flashing. The name of the data file is
saved and the unit is in the read mode again.
Is the file name shorter than 7 digits, a blank space must be set to each
remaining digit.
Starting / stopping measurement value display.
flashes during measurement data display (in the measurement mode
only).
Value
START
I
START
I
Press
key to start record: Data record is
running, display has changed to
STOP
and the down arrow
is blink-
ing.
STOP
I
Press
key to stop saving: Data record is
stopped, display has changed to
START
and the up arrow
is blinking.
The unit does not return automatically to the measurement mode, as
long as the arrows
or
in the display are blinking.
Press the
key to leave the editing mode. Then, after approx. 10 s,
the unit returns automatically to the measurement mode.
Deletion of all measurement data files (ending CSV) from USB memory stick.
Value
CLEAR
+
I
Press
keys at the same time to delete
files
CLEAR RUNNING
I
CSV files are being deleted
CLEAR DONE
I
CSV files have been deleted
Decimal separator
File name
Start / Stop
Deletion
Содержание AGC-100
Страница 2: ...Va AGC acuu C 100 um G Gaug ge Co Cop Ag ontro pyright 2016 gilent Inc oller ...
Страница 3: ......
Страница 24: ...Vacuum Gauge Controller AGC 100 24 Bargraph FSR ...
Страница 92: ...Vacuum Gauge Controller AGC 100 92 Notes ...
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...t qna96e1 ...