Vacuum Gauge Controller AGC-100
10
2 Technical
Data
Voltage
100 … 240 V (ac) ±10%
Frequency
50 … 60 Hz
Power consumption
≤
45 W
Overvoltage category
II
Protection class
1
Connection
European appliance connector IEC 320 C14
Temperature
Storage
Operation
–20 … +60 °C
+ 5 … +50 °C
Relative humidity
≤
80% up to +31 °C,
decreasing to 50% at +40 °C
Use indoors
only
max. altitude 2000 m NN
Pollution degree
II
Degree of protection
IP30
Number 1
Gauge connections per channel
RJ45 (FCC68), 8-pin (
→
19)
D-Sub, 15-pin, female (
→
19)
(connected in parallel)
Compatible gauges
Pirani PVG-500,
PVG-502, PVG-550, PVG-552
Pirani / Capacitance
PCG-750, PCG 752
Cold cathode
IMG-500
Cold cathode / Pirani
FRG-700, FRG-702
Hot ionization / Pirani
FRG-720, FRG-730
Capacitance CDG-500
Voltage
+24 V (dc) ±5%
Ripple <±1%
Current
0 … 1 A (per channel)
Power
25 W (per channel)
Fuse protection
1.5 A (per channel) with PTC element, self-
resetting after turning the unit off or disconnect-
ing the gauge. The supply conforms to the
grounded protective extra low voltage require-
ments.
Front panel
via 3 keys
Remote control
via USB type B interface
via Ethernet interface
Mains specifications
Ambience
Gauge connections
Gauge supply
Operation
Содержание AGC-100
Страница 2: ...Va AGC acuu C 100 um G Gaug ge Co Cop Ag ontro pyright 2016 gilent Inc oller ...
Страница 3: ......
Страница 24: ...Vacuum Gauge Controller AGC 100 24 Bargraph FSR ...
Страница 92: ...Vacuum Gauge Controller AGC 100 92 Notes ...
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...t qna96e1 ...