Vacuum Gauge Controller AGC-100
43
Transmission rate of the USB interface.
Value
BAUD USB
I
BAUD USB 9600
I
9600 baud
BAUD USB 19200
I
19200 baud
BAUD USB 38400
I
38400 baud
BAUD USB 57600
I
57600 baud
BAUD USB 115200
I
115200 baud (factory setting)
The display and setpoint adjustment range can be extended (the setting only
affects the control unit).
Use the parameter Pirani range extension only with Pirani and Pirani /
Capacitance gauges with display / measurement range up to
5×10
-5
mbar.
Available for the following gauges:
Pirani
(PVG)
Pirani / Capacitance
(PCG)
Cold cathode
(IMG)
Cold cathode / Pirani
(FRG-70x)
Hot ionization / Pirani
(FRG-720 / -730)
Capacitance
(CDG)
Value
RNG-EXT
I
RNG-EXT DISABLED
I
Deactivated (factory setting)
RNG-EXT ENABLED
I
Display extended to 5×10
-5
mbar
Switching behaviour of the error relay.
Value
ERR-RELAY
I
ERR-RELAY ALL
I
Switches for all errors (factory setting)
ERR-RELAY no SE
I
Only unit errors
ERR-RELAY CH 1
I
Error sensor 1 and unit error
Transmission rate
Pirani range extension
Error relay