Vacuum Gauge Controller AGC-100
30
Value
SP1-CH
I
Configuring a switching function.
SP1-CH DISABLED
I
Switching function 1 is factory-deactivated
SP1-CH ENABLED
I
Switching function 1 is always turned on
SP1-CH 1
I
Switching function 1 is assigned to
channel 1
The lower and the upper threshold of a switching function are always
assigned to the same channel. The last assignment is valid for both
thresholds.
Value
SP1-L
I
The lower threshold (Setpoint low) defines the
pressure at which the switching function is
activated when the pressure is dropping.
e.g.:
SP1-L 5.00-4
I
gauge dependent.
If another gauge type is connected, the
AGC-100 automatically adjusts the switching
threshold if required.
SPx-L min.
SPx-L max.
PVG5xx
I
2×10
-3 *)
= SPx-H max.
PCG75x
I
2×10
-3 *)
IMG500
I
1×10
-9
FRG70x
I
1×10
-9
FRG720
I
1×10
-8
FRG730
I
1×10
-8
CDG500
I
F.S. / 1000
all values in mbar, GAS=nitrogen
*)
2×10
-4
mbar if RNG-EXT (Pirani range extension) is activated (
→
43)
The minimum hysteresis between the upper and lower switching
threshold amounts to at least 10% of the lower threshold (logarithmic
gauges) or 1% of the full scale value (linear gauges). The upper
threshold is if necessary automatically adjusted to a minimum
hysteresis. This prevents unstable states.
Configuring a switching
function
Limits of the lower switching
thresholds
Содержание AGC-100
Страница 2: ...Va AGC acuu C 100 um G Gaug ge Co Cop Ag ontro pyright 2016 gilent Inc oller ...
Страница 3: ......
Страница 24: ...Vacuum Gauge Controller AGC 100 24 Bargraph FSR ...
Страница 92: ...Vacuum Gauge Controller AGC 100 92 Notes ...
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...t qna96e1 ...