Vacuum Gauge Controller AGC-100
31
Value
SP1-H
I
The upper switching threshold (Setpoint high)
defines the pressure at which the switching
function is deactivated when the pressure is
rising.
e.g.:
SP1-H 1500
I
gauge dependent.
If another gauge type is connected, the
AGC-100 automatically adjusts the threshold
if required.
SPx-H min.
SPx-H max.
PVG5xx
I
= SPx-L min.
1×10
3
PCG75x
I
1.5×10
3
IMG500
I
1×10
-2
FRG70x
I
1×10
3
FRG720
I
1×10
3
FRG730
I
1×10
3
CDG500
I
F.S.
all values in mbar, GAS=nitrogen
The minimum hysteresis between the upper and lower switching
threshold amounts to at least 10% of the lower threshold (logarithmic
gauges) or 1% of the full scale value (linear gauges). This prevents
unstable states.
Limits of the upper switching
thresholds