Vacuum Gauge Controller AGC-100
48
Behaviour of the system control (watchdog control) in the event of an error.
Setting
WATCHDOG
I
WATCHDOG AUTO
I
The system automatically acknowledges
error messages of the watchdog after 2 s
(factory setting)
WATCHDOG OFF
I
Error messages of the watchdog have to be
acknowledged by the operator
The measurement units Torr and Micron can be suppressed in the corresponding
parameter setting
UNIT TORR
(
→
42).
Setting
TORR-LOCK
I
TORR-LOCK OFF
I
Measurement units Torr and Micron
available (factory setting)
TORR-LOCK ON
I
Measurement units Torr and Micron not
available
The keylock function prevents inadvertent entries in the parameter mode and thus
malfunctions.
Setting
KEY-LOCK
I
KEY-LOCK OFF
I
Keylock function disabled (factory setting)
KEY-LOCK ON
I
Keylock function enabled
Test of the program memory.
Test sequence
FLASH
+
I
Press
keys at the same time to start test
FLASH RUN
I
Test in progress (very briefly)
FLASH PASS
I
Test completed, no error found. After the
test, an 8-digit checksum
(e.g.
FLASH 0x12345678
) is displayed.
FLASH ERROR
I
Test completed, error found. After the test,
an 8-digit checksum
(e.g.
FLASH 0x12345678
) is displayed.
If the error persists after repeating the test,
please contact your nearest Pfeiffer Vacuum
service center.
Watchdog control
Torr lock
Keylock
FLASH test
Содержание AGC-100
Страница 2: ...Va AGC acuu C 100 um G Gaug ge Co Cop Ag ontro pyright 2016 gilent Inc oller ...
Страница 3: ......
Страница 24: ...Vacuum Gauge Controller AGC 100 24 Bargraph FSR ...
Страница 92: ...Vacuum Gauge Controller AGC 100 92 Notes ...
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ...t qna96e1 ...