7-12
Specifications and Regulatory Information
Specifications
Table 7-14. Option 004 EELED Sources
a. Typical
Minimum Output Power Spectral Density
c
d
50/125
µm
fiber
62.5/125
µm
fiber
–50 dBm/nm (10 nW/nm)
–46 dBm/nm (25 nW/nm)
Output Stability
b
±
0.02 dB over 10 minutes
Lamp Lifetime, Mean Time Between Failures
>5000 hours
a. Filtered below 850 nm.
b. With applied input fiber 9/125 µm.
c. Typical
d. Includes power in full numerical aperture of fiber.
Agilent 86142B, 86146B only
Agilent 86142B, 86146B only
Minimum Spectral Power Density
1300 to 1320 nm, 1540 to 1560 nm - option 004
1250 to 1620 nm - option 004
a
> –40 dBm/nm (100 nW/nm)
> –60 dBm/nm (1 nW/nm)
Return Loss
a
With straight connector
>25 dB
Stability (ambient temp <±1
°
C)
a
Over 15 minutes
Over 6 hours
<
±
0.02 dB
<
±
0.05 dB
Содержание 8614 B Series
Страница 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Страница 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Страница 13: ...1 5 Getting Started Product Overview ...
Страница 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 16: ...1 8 Getting Started Setting Up the Analyzer Setting Up the Analyzer Step 1 Receive and Inspect the Shipment ...
Страница 26: ...1 18 Getting Started The Softkey Panels ...
Страница 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 40: ...1 32 Getting Started Product Options and Accessories ...
Страница 75: ...3 Function Reference ...
Страница 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Страница 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 264: ...7 16 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 271: ......