3-37
Function Reference
Marker Setup
Peak Excursion
Determines (in dB) which side modes are included in the measurements. To be accepted, each
trace peak must rise, and then fall, by at least the peak excursion value about a given spectral
component. The default value is 3 dB. Setting the value too high may result in not identifying a
side mode. Setting the value too low may cause unwanted responses, including noise spikes, to
be identified. Use the Marker Setup panel to change the peak excursion value.
Changing the peak excursion. The peak searches may not recognize valid signals near the noise
floor when the peak excursion definition is less than 3 dB. Thus, before performing peak searches
on signals near the noise floor, reduce the peak excursion value. However, reducing the excur-
sion value often causes peak searches to identify noise spikes. To avoid this misidentification, use
video averaging or video filtering to reduce the noise floor variance to a value less than the peak
excursion definition.
Pit Excursion
Sets the pit excursion value for the marker search routines.
The pit excursion value is used to determine whether or not a local minimum in the trace is to be
considered a pit. To qualify as a pit, both sides of the local minimum must rise by at least the pit
excursion value.
Changing the pit excursion. The pit searches may not recognize valid signals near the noise floor
when the pit excursion definition is less than 3 dB. Thus, before performing pit searches on sig-
nals near the noise floor, reduce the pit excursion value. However, reducing the excursion value
Peak excursion criteria
Pit excursion criteria
Содержание 8614 B Series
Страница 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Страница 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Страница 13: ...1 5 Getting Started Product Overview ...
Страница 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 16: ...1 8 Getting Started Setting Up the Analyzer Setting Up the Analyzer Step 1 Receive and Inspect the Shipment ...
Страница 26: ...1 18 Getting Started The Softkey Panels ...
Страница 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 40: ...1 32 Getting Started Product Options and Accessories ...
Страница 75: ...3 Function Reference ...
Страница 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Страница 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 264: ...7 16 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 271: ......