3-81
Function Reference
Sweep Points
Remote Commands
SENSe:WAVelength:STOP
Sweep Points
Specifies the number of data points taken for a sweep. The more data points the better the trace
resolution, but the longer the sweep time. You can select from 3 to 10001 points. Enter the num-
ber of data points using the step keys, numeric key pad, or knob. The default is 1001 points.
Key Path
Traces > Trace Setup > Sweep Points
Remote Commands
SENSe:SWEep:POINts
Sweep Time
Specifies the amount of time required for the instrument to sweep the current measurement
range. The instrument automatically selects sweep times based on coupling of the following
instrument settings:
• wavelength span
• resolution bandwidth
• video bandwidth
• sensitivity
• trace length
• power level
Coupling of these parameters yields optimum amplitude accuracy. When Sweep Time is set to
Auto, the instrument always uses the fastest sweep possible while still maintaining the specified
accuracy. Coupled, sweep times range from 56.3 mn to a maximum value that depends on the
number of trace points used to draw the trace. This relationship is shown in the following equa-
tion:
The default number of trace points is 1001, so the maximum sweep time is normally 100 sec-
onds. When Sweep Time is in manual mode, the sweep time can be set from 56.3 ms to a maxi-
mum of 1000 seconds. If you change the number of trace points, the maximum sweep time
changes as well.
56,3ms
sweep time
1min
(
)
trace points
(
)
≤
≤
Содержание 8614 B Series
Страница 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Страница 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Страница 13: ...1 5 Getting Started Product Overview ...
Страница 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 16: ...1 8 Getting Started Setting Up the Analyzer Setting Up the Analyzer Step 1 Receive and Inspect the Shipment ...
Страница 26: ...1 18 Getting Started The Softkey Panels ...
Страница 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 40: ...1 32 Getting Started Product Options and Accessories ...
Страница 75: ...3 Function Reference ...
Страница 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Страница 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 264: ...7 16 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 271: ......