3-46
Function Reference
OSNR Marker
N o t e
For Agilent 86146B only, if filter mode is enabled, the noise marker, marker BW, delta marker, and
OSNR marker are disabled.
= noise marker symbols
N = interpolated noise
N
a
= interpolated noise amplitude
P = peak
P
a
= peak amplitude
P
w
= center wavelength
Center amplitude = P
a
-N
a
OSNR = Center amplitude / N
a
(normalized)
Interpolating Noise
To accurately measure noise, the noise level must be determined at the signal wavelength. The
measurement cannot be directly performed because the signal power level masks the noise. To
estimate the noise at the signal wavelength, the OSA measures the noise power above and
below the channel wavelength at the predetermined offset or pits (depending on the Noise
Method selected in the Marker Setup panel). Next, the values are linearly interpolated to deter-
mine the noise value at the channel wavelength.
Содержание 8614 B Series
Страница 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Страница 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Страница 13: ...1 5 Getting Started Product Overview ...
Страница 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 16: ...1 8 Getting Started Setting Up the Analyzer Setting Up the Analyzer Step 1 Receive and Inspect the Shipment ...
Страница 26: ...1 18 Getting Started The Softkey Panels ...
Страница 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 40: ...1 32 Getting Started Product Options and Accessories ...
Страница 75: ...3 Function Reference ...
Страница 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Страница 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 264: ...7 16 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 271: ......