6-6
Maintenance
Printer Head Cleaning Procedure
5 Unscrew the retaining screw that secures the sheet-metal cover that protects the printer head
from electrostatic discharge. Slide the sheet-metal cover towards the retaining screw and then lift
it straight up to remove.
6 Lift the printer head lever to the vertical position. Then, tilt the lever towards the instrument’s rear
panel to rotate the printer head up.
7 Clean the printer head using a cotton swab and isopropyl alcohol.
C A U T I O N
Use of other cleaning materials or fluids may damage the printer.
8 After the printer head has thoroughly dried, use the printer head lever to return the printer head
to its original position.
Содержание 8614 B Series
Страница 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Страница 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Страница 13: ...1 5 Getting Started Product Overview ...
Страница 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 16: ...1 8 Getting Started Setting Up the Analyzer Setting Up the Analyzer Step 1 Receive and Inspect the Shipment ...
Страница 26: ...1 18 Getting Started The Softkey Panels ...
Страница 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 40: ...1 32 Getting Started Product Options and Accessories ...
Страница 75: ...3 Function Reference ...
Страница 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Страница 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 264: ...7 16 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 271: ......