3-45
Function Reference
Normal Marker Units (Marker Setup Panel)
Related Functions
Active Marker, Delta Marker
Remote Commands
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:INTerpolation
Normal Marker Units (Marker Setup Panel)
Sets the marker X-axis readout for frequency or wavelength when the instrument is in a non-zero
span. The X-axis default units is nm (nanometers). This setting controls only the normal marker X-
axis readout and the delta reference readout. This setting controls all four normal markers. The
bandwidth and delta offset markers have their own setting.
Key Path
Markers > Marker Setup > Normal Marker Units
Related Functions
Active Marker
Remote Commands
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:X:READout
OSNR Marker
Indicates the signal quality based on the signal strength and noise level. Turning on the OSNR
marker will display four markers: an active marker, a center marker, and two noise markers. To
adjust the active marker to the desired measurement location, use the step keys, knob, or key-
board. The center marker is positioned at the closest peak to the active marker. The location of
the noise markers is determined by the selection of the Noise Method and the center marker. The
noise measurement is referenced to 0.1 nm noise resolution bandwidth.
Содержание 8614 B Series
Страница 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Страница 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Страница 13: ...1 5 Getting Started Product Overview ...
Страница 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 16: ...1 8 Getting Started Setting Up the Analyzer Setting Up the Analyzer Step 1 Receive and Inspect the Shipment ...
Страница 26: ...1 18 Getting Started The Softkey Panels ...
Страница 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 40: ...1 32 Getting Started Product Options and Accessories ...
Страница 75: ...3 Function Reference ...
Страница 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Страница 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 264: ...7 16 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 271: ......