5-31
Status Listings
Application-Specific Warnings
8014
desc = “Print statement ignored: no path is specified”
help = “The PRINT statement needs to be after a PATH statement to indicate which
PATH data is to be printed. The PRINT statement at the line number specified came
before any PATH statement. This PRINT statement will not trigger any printouts.”
8015
desc = “DEFAULT cannot be used for a parameter.”
help = “A parameter cannot be specified as DEFAULT. The parameter needs to be
either a previously defined variable name or an actual value. The bad parameter
was found on the line number specified. The number in the parentheses is the
number of the parameter causing the problem.”
8016
desc = “A duplicate command is in the spec file.”
help = “Certain commands should be used only once in the specification file. The
duplicate command was found on the line number specified. The command in the
parentheses is the duplicate.”
8017
desc = “A required command is missing from the spec file.”
help = “Certain commands must be used in the specification file. The missing
command is listed in the parentheses. This command should be used only once in
the file.”
8018
desc = “The normalization interval limited to maximum.”
help = “The normalization interval has a maximum value of 24 and a minimum
value of 0.1 hours. The interval specified in the specification file is outside this
range. The specification file can still be used but the calibration interval will be 24
hours.”
8019
desc = “The spec file could not be loaded.”
help = “The application found a problem with the specification file. Possible causes
can include the following: the file is incompatible with the application the file is
from a previous version of the application the file was corrupted. The file with the
problem is specified within the parentheses. Try re-importing the specification file.”
8020
desc = “The default spec file was loaded.”
help = “The default specification file for the application was loaded. This is done
when the application is first started or if a problem occurred when trying to load
another specification file. Refer to the previous warnings for information on any
problems loading another file.”
8021
desc = “The spec file could not be found.”
help = “The application tried to load a specification file which could not be found in
the internal memory. This may have occurred if the last file used by the application
was deleted. The filename is specified within the parentheses.”
Table 5-7. Application-Specific Warnings (3 of 6)
Error Number
Error Description [description/explanation/examples]
Содержание 8614 B Series
Страница 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Страница 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Страница 13: ...1 5 Getting Started Product Overview ...
Страница 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 16: ...1 8 Getting Started Setting Up the Analyzer Setting Up the Analyzer Step 1 Receive and Inspect the Shipment ...
Страница 26: ...1 18 Getting Started The Softkey Panels ...
Страница 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 40: ...1 32 Getting Started Product Options and Accessories ...
Страница 75: ...3 Function Reference ...
Страница 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Страница 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 264: ...7 16 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 271: ......