5-19
Status Listings
OSA Warnings
5013
desc = “The floppy disk is write protected”
help = “An operation was requested which writes to the floppy disk. The disk
currently in the drive is write-protected. Turn off write protection on the current
floppy disk or replace it with another disk.”
5014
desc = “The internal memory is full”
help = “An operation was requested which uses internal memory. There is not
enough free space left in internal memory for this operation. You must make room
by deleting one or more files.”
5015
desc = “Up/down keys will not alter the span when it is zero”
help = “The up/down step keys may not be used to alter the span setting when it is
zero. A non-zero numeric value must be entered for span in order to leave zero span
mode.”
5016
desc = “Incompatible measurement file”
help = “An attempt was made to recall a measurement file which is incompatible
with the current version of software.”
5017
desc = “The requested file does not exist”
help = “An attempt was made to access a file which does not exist. Check the
spelling of the file name against the listing of available files.”
5020–5030
desc = “I/O Error”
help “An unexpected error occurred during the I/O operation. Please try the
operation again. If the operation involves the floppy disk drive, try a different floppy
disk. If the error persists, please make a note of the error number and contact the
nearest Agilent Technologies Instrument support center for assistance. In the U.S.,
call (800) 403-0801. See the
Agilent 86140B Series Users Guide
for a listing of the
Agilent sales and service offices.
5031
desc = “Could not initialize floppy”
help => “A request was made to initialize a floppy disk. The operation did not
succeed. Check to see that there is a floppy disk inserted in the drive and that the
disk is not write protected. This can also be caused by a defective floppy disk.”
5032–5043
desc = “I/O Error”
help “An unexpected error occurred during the I/O operation. Please try the
operation again. If the operation involves the floppy disk drive, try a different floppy
disk. If the error persists, please make a note of the error number and contact the
nearest Agilent Technologies Instrument support center for assistance. In the U.S.,
call (800) 403-0801. See the
Agilent 86140B Series Users Guide
for a listing of the
Agilent sales and service offices.
Table 5-6. OSA Warnings (3 of 12)
Error Number
Error Description [description/explanation/examples]
Содержание 8614 B Series
Страница 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Страница 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Страница 13: ...1 5 Getting Started Product Overview ...
Страница 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 16: ...1 8 Getting Started Setting Up the Analyzer Setting Up the Analyzer Step 1 Receive and Inspect the Shipment ...
Страница 26: ...1 18 Getting Started The Softkey Panels ...
Страница 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 40: ...1 32 Getting Started Product Options and Accessories ...
Страница 75: ...3 Function Reference ...
Страница 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Страница 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 264: ...7 16 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 271: ......