1-12
Getting Started
Making a Measurement
Making a Measurement
This procedure will introduce you to the Agilent 8614x series optical spectrum analyzer front
panel controls. By following this procedure you will do the following:
• Perform an auto alignment
• Perform a peak search
• Use a delta marker
• Print the display
Refer to “The Menu Bar” on page 1-15
and
“The Softkey Panels” on page 1-16
.
Instrument setup
A source signal must be present at the input of the optical spectrum analyzer. In this procedure a
Fabry-Perot laser is used as the source. You can use another source or the optional 1310/1550
nm eeled. If another source is being used, the display will differ from those shown.
To set the OSA to a known state
• Press the front-panel Preset key to set the instrument to a known state. For a complete descrip-
tion of preset conditions, see
page 3-59
.
Содержание 8614 B Series
Страница 1: ...Agilent 8614xB Series Optical Spectrum Analyzer User s Guide ...
Страница 12: ...1 4 Getting Started Product Overview Agilent 8614xB Front and Rear Panels ...
Страница 13: ...1 5 Getting Started Product Overview ...
Страница 15: ...1 7 Getting Started Product Overview Figure 1 2 Display Annotations ...
Страница 16: ...1 8 Getting Started Setting Up the Analyzer Setting Up the Analyzer Step 1 Receive and Inspect the Shipment ...
Страница 26: ...1 18 Getting Started The Softkey Panels ...
Страница 31: ...1 23 Getting Started The Softkey Panels The Systems Menus continued ...
Страница 40: ...1 32 Getting Started Product Options and Accessories ...
Страница 75: ...3 Function Reference ...
Страница 186: ...4 16 Remote Front Panel Operation Remote Front Panel ...
Страница 226: ...6 2 Maintenance Changing the Printer Paper Changing the Printer Paper ...
Страница 264: ...7 16 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 271: ......