![Agilent Technologies 5975C TAD VL MSD Скачать руководство пользователя страница 169](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605169.webp)
CI Maintenance
5
5975 Series MSD Troubleshooting and Maintenance Manual
169
To Refill the CI Calibrant Vial
Materials needed
•
PFDTD calibrant (8500-8510)
Procedure
1
Set the reagent gas flow to
Gas Off.
2
Vent the MSD.
3
Remove the capillary column from the GC/MSD interface.
4
Pull the MSD away from the GC to expose the calibration vial and valve.
See
5
Loosen the collar holding the calibration vial in place. Do not remove the
collar.
6
Remove the calibrant vial. See
7
Fill the vial no higher than the bottom of the internal tube with fresh
PFDTD calibrant (8500-8510).
8
Replace the vial and tighten the collar.
9
Reposition the MSD next to the GC. See
.
10
Reinstall the capillary column.
11
Pump down the MSD.
12
Purge the calibration valve. See
C A U T I O N
Do
not
rinse the vial with any solvents.
Never
expose the inside of the vial to
chlorinated solvents or isopropyl alcohol or water — this will result in severe loss of CI
sensitivity.
C A U T I O N
After removing the calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in severe contamination of the ion source and damage to the filament and
electron multiplier.
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...