![Agilent Technologies 5975C TAD VL MSD Скачать руководство пользователя страница 105](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605105.webp)
General Maintenance
4
5975 Series MSD Troubleshooting and Maintenance Manual
105
To Reconnect the MSD to the 7890A GC/7820A/6890 GC
Materials needed
•
Wrench, open-end, 1/4-inch
×
5/16-inch
(8710-0510)
Procedure
1
Position the MSD so the end of the GC/MSD interface is near the GC
(
2
Reconnect the GC/MSD interface cable.
3
Slide the MSD to its regular position next to the GC.
Be careful not to damage the GC/MSD interface as it passes into the GC.
Make sure the end of the GC/MSD interface extends into the GC oven.
4
The foreline pump may be located on the floor, on the lab bench next to or
behind the MSD, or under the analyzer chamber at the back of the MSD.
5
Reinstall the capillary column.
6
Pump down the MSD.
7
Turn on the GC. Enter appropriate temperature setpoints for the GC/MSD
interface and GC oven.
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...