General Troubleshooting
2
5975 Series MSD Troubleshooting and Maintenance Manual
39
High abundances at
m/z
18, 28, 32, and 44 or at
m/z
14 and 16
•
System was recently vented (residual air and water)
•
Air leak. Large peaks at
m/z
14 and 16 are symptomatic of especially large
leaks.
Mass assignments are incorrect
Small shape changes at the top of the mass peaks can cause 0.1
m/z
shifts in
mass assignments. Shifts greater than 0.2
m/z
indicate a possible malfunction.
•
MSD has not had enough time to reach thermal equilibrium
•
Large variations in the temperature of the laboratory
•
MSD has not been tuned recently, or at the temperature at which it is
operating
•
Incorrect tune file (inappropriate parameters)
Peaks have precursors
The tune report lists the size of the precursors for the tune masses. Small
precursors are not unusual. If the precursors are unacceptably large for your
application, one of the following may be responsible:
•
Repeller voltage is too high
•
Peaks are too wide
•
Incorrect DC polarity on the quadrupole mass filter
•
Dirty quadrupole mass filter
Peak widths are inconsistent
•
MSD has not had enough time to reach thermal equilibrium
•
Large variations in the temperature of the laboratory
•
Incorrect tuning
•
Calibration vial(s) empty or almost empty
•
Calibration valve(s) not working correctly
•
Dirty ion source
•
Electron multiplier is nearing the end of it useful lifetime
•
Ground loop problems
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...