General Troubleshooting
2
5975 Series MSD Troubleshooting and Maintenance Manual
37
Poor sensitivity
•
Incorrect tuning
•
Tune file that does not match the type of analysis
•
Repeller voltage is too low
•
Incorrect temperatures (oven, GC/MSD interface, ion source, or mass filter)
•
Incorrect sample concentration
•
Leaking GC inlet*
•
Dirty GC inlet
•
Incorrect split ratio
•
Purge off time in splitless mode is too short
•
Excessive pressure in the MSD
•
Dirty ion source
•
Air leak
•
Poor filament operation
•
Detector (HED electron multiplier) is not working correctly
•
Incorrect mass filter polarity
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Poor repeatability
•
Dirty syringe needle
•
Dirty GC inlet
•
Leaking GC inlet*
•
Injection is too large
•
Loose column connections
•
Variations in pressure, column flow, and temperature
•
Dirty ion source
•
Loose connections in the analyzer
•
Ground loops
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...