![Agilent Technologies 5975C TAD VL MSD Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605024.webp)
24
5975 Series MSD Troubleshooting and Maintenance Manual
1
Introduction
Operating precautions
•
Turn off the hydrogen at its source every time you shut down the GC or
MSD.
•
Turn off the hydrogen at its source every time you vent the MSD (do not
heat the capillary column without carrier gas flow).
•
Turn off the hydrogen at its source every time shutoff valves in an MSD are
closed (do not heat the capillary column without carrier gas flow).
•
Turn off the hydrogen at its source if a power failure occurs.
•
If a power failure occurs while the GC/MSD system is unattended, even if
the system has restarted by itself:
1
Immediately turn off the hydrogen at its source.
2
Turn off the GC.
3
Turn off the MSD and allow it to cool for 1 hour.
4
Eliminate
all
potential sources of ignition in the room.
5
Open the vacuum manifold of the MSD to atmosphere.
6
Wait at least 10 minutes to allow any hydrogen to dissipate.
7
Start up the GC and MSD as normal.
When using hydrogen gas
check the system for leaks to prevent possible fire
and explosion hazards based on local Environmental Health and Safety (EHS)
requirements. Always check for leaks after changing a tank or servicing the
gas lines. Always make sure the vent line is vented into a fume hood.
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...