General Troubleshooting
•
Electrical leads to the detector are not connected correctly
•
HED power supply output cable failed
•
Electrical leads to the ion source are not connected correctly
•
Filament shorted to the source body
Temperature control disabled
•
One of the heater fuses has failed
•
MSD electronics are not working correctly
Temperature control fault
This indicates that something has gone wrong with the temperature control of
either the ion source or mass filter (quad) heater. The cause can be further
isolated by selecting
Status/MS Temp Ctlr Status
in the Tune and Vacuum
Control view. One of the following should be displayed as the cause:
•
Source temperature sensor is open
•
Source temperature sensor is shorted
•
Mass filter (quad) temperature sensor is open
•
Mass filter (quad) temperature sensor is shorted
•
No heater voltage (heater fuse has probably failed)
•
Heater voltage is too low
•
Temperature zone has timed out (heater failed, bad heater wiring, or loose
temperature sensor)
•
Problem with the temperature control electronics
•
Source heater is open
•
Source heater is shorted
•
Mass filter heater is open
•
Mass filter heater is shorted
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...