Compression seals
A compression fitting consists of a threaded fitting on the analyzer chamber
and a threaded collar with a ferrule and O-ring. A cylindrical part fits inside
the collar. Tightening the collar presses the ferrule, compressing the O-ring
around the part. The calibration vials use compression seals.
High voltage feedthrough seal
The high voltage (HED) feedthrough seal is an O-ring that is compressed
against the side plate by a threaded collar.
Figure 38
Vacuum seals
Side plate O-ring seal
KF seal with internal
centering ring
KF seal with external
centering ring
Compression seal
(clamp not shown)
(clamp not shown)
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...