General Symptoms
This section describes symptoms you might observe when first turning on the
GC/MSD system. All of these symptoms would prevent operation of the
system.
GC does not turn on
Nothing happens when the GC is switched on. The GC fans do not turn on and
the keypad display does not light.
•
Disconnected GC power cord
•
No voltage or incorrect voltage at the electrical outlet
•
Failed fuse in the GC
•
GC power supply is not working correctly
MSD does not turn on
Nothing happens when the MSD is switched on. The foreline pump does not
start. The cooling fan for the high vacuum pump does not turn on. The local
control panel is not on.
•
Disconnected MSD power cord
•
No voltage or incorrect voltage at the electrical outlet
•
Failed primary fuses
•
MSD electronics are not working correctly
Foreline pump is not operating
The MSD is receiving power (the fan is operating and the local control panel is
lit) but the foreline pump is not operating.
•
Large air leak (usually the analyzer door open) has caused pumpdown
failure. See
174. You must power
cycle the MSD to recover from this state.
•
Disconnected foreline pump power cord
•
Malfunctioning foreline pump
•
Check power switch on foreline pump
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...