![Agilent Technologies 5975C TAD VL MSD Скачать руководство пользователя страница 159](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605159.webp)
General Maintenance
4
5975 Series MSD Troubleshooting and Maintenance Manual
159
To adjust the quad frequency
Materials needed
•
Screwdriver, flat-blade, large (8730-0002)
Procedure
1
Make sure the MSD is at thermal equilibrium. It takes at least 2
hours
after
all heated zones have reached their setpoints for the MSD to reach
thermal equilibrium.
2
Open the analyzer cover.
3
Make sure the RF cover on the side board is secure and no screws are
missing. A loose RF cover or missing screw can
significantly
affect coil
adjustment.
4
In the Tune and Vacuum Control view, select
Optimize Quadrupole Frequency
from the Execute menu
.
5
Enter an
m/z
value of 100.
6
Slowly turn the quad frequency adjustment screws to minimize the voltage
displayed (
Turn the adjustment screws alternately. Turn each screw only a little bit at
a time. Keep the screws at
equal
extension. The minimum voltage is
typically between 50 and 70
mV.
7
When the voltage is minimized, click
Stop
.
WA R N I N G
Do not remove any other covers. Dangerous voltages are present under these
covers.
C A U T I O N
Do not use a coin to adjust the screws. If you drop it, it could fall into the electronics
fan and cause significant damage.
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...