![Agilent Technologies 5975C TAD VL MSD Скачать руководство пользователя страница 113](http://html1.mh-extra.com/html/agilent-technologies/5975c-tad-vl-msd/5975c-tad-vl-msd_troubleshooting-and-maintenance-manual_2867605113.webp)
General Maintenance
4
5975 Series MSD Troubleshooting and Maintenance Manual
113
To Reinstall the EI Calibration and Vent Valve Assembly
Materials needed
•
Calibration valve (G3170-60204)
•
O-ring for calibration valve (0905-1217)
•
PFTBA (05971-60571) or other tuning compound
•
Screwdriver, Torx T-15 (8710-1622)
Procedure
1
Remove the old valve assembly. See
.
2
Be sure the valve O-ring is in place. If it is worn or damaged, replace it.
3
Install the calibration and vent valve assembly and tighten the screws that
hold it in place.
4
Reconnect the calibration valve cable to the connector next to the fan.
5
Remove the vial from the new calibration valve. See
. The valve is
supplied with a vial already installed.
6
Fill and reinstall the calibration vial. See
7
Pump down the MSD.
8
Select
Purge Calibrant Valve
from the Vacuum menu in the Tune and Vacuum
Control view.
C A U T I O N
Failure to purge the calibration valve will damage the filaments and detector.
Содержание 5975C TAD VL MSD
Страница 1: ...Agilent Technologies Agilent 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 8: ...8 5975 Series MSD Troubleshooting and Maintenance Manual...
Страница 56: ...56 5975 Series MSD Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Страница 82: ...82 5975 Series MSD Troubleshooting and Maintenance Manual 3 CI Troubleshooting...
Страница 196: ...196 5975 Series MSD Troubleshooting and Maintenance Manual 6 Vacuum System...
Страница 230: ...230 5975 Series MSD Troubleshooting and Maintenance Manual 8 Electronics...
Страница 249: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 249 Figure 59 Turbo pump and related parts 1 2 3 4 5...
Страница 253: ...Parts 9 5975 Series MSD Troubleshooting and Maintenance Manual 253 Figure 61 Analyzer parts 7 under clips 3 4 5 6...
Страница 256: ...256 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts Figure 62 EI source 1 9 10 12 11 4 8 5 3 3 7 7 2...
Страница 270: ...270 5975 Series MSD Troubleshooting and Maintenance Manual 9 Parts...